Jokoh EX-D Electrolyte Analyzer
Features
• Measurement Parameter : Na,+ K,+ Cl-
• Measurement Sample : Whole Blood, Plasma, Serum, Diluted Urine
• Measurement Speed : 300 tests/hour
• Newly developed high sensitive electrode allows only one time calibration in a day to obtain
• very accurate result of C.V. ≤ 0.5%
×
Request a Quote
Jokoh electrolyte analyzer
- Jokoh ex-d is high sensitivity electrolyte analyzer with auto-sampler(20 Samples)
- EX-D electrolyte analyzer are used to measure sodium, potassium, and chloride in whole blood, serum, or plasma.
- Ion Selective Electrode (ISE) and Flame Emission Photometry (FEP) are the two most popular analytical techniques. Majority of analyzers measure electrolyte using ISE technology.
Technical Specification
Measurement items |
Na+, K+and Cl- simultaneously measured |
Measurement range |
|
Serum |
|
Na+ | 70.0~200.0 [mmol/L] |
K+ | 1.00~20.00 [mmol/L] |
Cl- | 70.0~200.0 [mmol/L] |
Urine |
|
Na+ | 10.0~500.0 [mmol/L] |
K+ | 1.00~200.0 [mmol/L] |
Cl- | 10.0~550.0 [mmol/L] |
Adjusted dialysate |
|
Na+ | 130.0~150.0 [mmol/L] |
K+ | 1.00~3.00 [mmol/L] |
Cl- | 100.0~120.0 [mmol/L] |
Dialysate A |
|
Na+ | 100.0~120.0 [mmol/L] |
K+ | 1.00~3.00 [mmol/L] |
Cl- | 100.0~120.0 [mmol/L] |
Dialysate B |
|
Na+ | 20.0~40.0 [mmol/L] |
Measurement pH range |
7.0~8.0 (no change in pH during measurement) |
Measurement principle |
Ion Selective Electrode method |
Na+ | Crown ether solution membrane electrode |
K+ | Valinomycin solution membrane electrode |
Cl- | Grade-4 ammonium salt solution membrane electrode |
Measurement targets |
Whole blood, serum, plasma, urine (to be diluted), adjusted dialysate,dialysate A, dialysate B |
Processing capacity |
100 samples/hour (max.) |
Sample aspiration volume |
95 [μL] |
300 [μL] in “urine” mode | |
60 [μL] or more in small volume sample measurement | |
Sample table |
|
Routine measurement | 20 samples |
Urgent sample | 1 sample |